    <ArticleTitle>Comparison of cyclic fatigue resistance of XP-endo Shaper, HyFlex CM, FlexMaster and Race instruments</ArticleTitle>
    <FirstPage>208</FirstPage>
    <LastPage>212</LastPage>
    <ELocationID EIdType="doi">10.15171/joddd.2018.032</ELocationID>
    <Language>EN</Language>
    <AuthorList>
      <Author>
        <FirstName>Mehmet</FirstName>
        <LastName>Adiguzel</LastName>
      </Author>
      <Author>
        <FirstName>Ipek</FirstName>
        <LastName>Isken</LastName>
      </Author>
      <Author>
        <FirstName>Ismail Ilker</FirstName>
        <LastName>Pamukcu</LastName>
      </Author>
    </AuthorList>
    <PublicationType>Journal Article</PublicationType>
    <ArticleIdList>
      <ArticleId IdType="doi">10.15171/joddd.2018.032</ArticleId>
    </ArticleIdList>
    <History>
      <PubDate PubStatus="received">
        <Year>2018</Year>
        <Month>05</Month>
        <Day>01</Day>
      </PubDate>
      <PubDate PubStatus="accepted">
        <Year>2018</Year>
        <Month>07</Month>
        <Day>28</Day>
      </PubDate>
    </History>
    <Abstract>Background. The aim of this study was to compare the cyclic fatigue resistance of XP-endo Shaper, HyFlex CM, FlexMaster and Race rotary instruments at body temperature (37±1°C). Methods. Twenty XP-endo Shaper (#30/.01), 20 HyFlex CM (#30/.04), 20 FlexMaster (#30/.04) and 20 Race (#30/.04) instruments were tested at body temperature (n=20). The instruments were evaluated in artificial canals with a 3-mm radius of curvature and 60° angle of curvature to the center of the 1.5-mm-wide canal. Each instrument was rotated until fracture occurred and the number of cycles to failure (NCF) recorded. Data were analyzed using one-way ANOVA and Tukey HSD tests (P&lt;0.05). Results. The difference in the NCF of all the instruments was statistically significant (P&lt;0.05). The order of the instruments from the highest to the lowest NCF was as follows: XP-endo Shaper (3064.0±248.1), HyFlex CM (1120.5±106.1), FlexMaster (569.8±48.4) and Race (445.5±53.5). Conclusion. Under the limitations of the present study, XP-endo Shaper instruments were more resistant to cyclic fatigue than the #30/.04 nickel-titanium rotary instruments immersed in water at simulated body temperature.</Abstract>
    <ObjectList>
      <Object Type="keyword">
        <Param Name="value">Body temperature</Param>
      </Object>
      <Object Type="keyword">
        <Param Name="value">cyclic fatigue</Param>
      </Object>
      <Object Type="keyword">
        <Param Name="value">FlexMaster</Param>
      </Object>
      <Object Type="keyword">
        <Param Name="value">HyFlex CM</Param>
      </Object>
      <Object Type="keyword">
        <Param Name="value">Race</Param>
      </Object>
      <Object Type="keyword">
        <Param Name="value">XP-endo Shaper</Param>