Comparison of the accuracy of conventional and digital radiography in root canal working length determination: An in vitro study

J Dent Res Dent Clin Dent Prospects, 11(3), 161-165; DOI:10.15171/joddd.2017.029

Clinical Dentistry

Comparison of the accuracy of conventional and digital radiography in root canal working length determination: An in vitro study

Mohammad Mahdi Yaghooti Khorasani1, Hamed Ebrahimnejad2*

1 Department of Endodontics, School of Dentistry, Rafsanjan University of Medical Sciences, Rafsanjan, Iran
2 Department of Oral and Maxillofacial Radiology, School of Dentistry, Kerman University of Medical Sciences, Kerman, Iran

*Corresponding Author; Email: hsimple11@gmail.com

© 2017 by Tabriz University of Medical Sciences

Abstract

Background. Digital radiography has widespread use in endodontics. Determining a correct working length is vital for a proper endodontic therapy. The aim of this study was to compare the accuracy of conventional and digital radiographic techniques for root canal working length determination.

Methods. After determining the real working lengths of 50 permanent maxillary central incisors (gold standard), the conventional (E- and F-speed films) and digital (CCD, PSP) images were obtained using the parallel technique. The mean registered working length of each modality was compared with the other and with the gold standard using one-way ANOVA at P<0.05.

Results. NNo significant difference was found between the recorded working length values using the conventional and digital radiographic techniques (P=0.828).

Conclusion.Within the limitations of this study, it was concluded that there was no difference between the measurement accuracy of CCD, PSP and conventional imaging techniques in root canal working length determination.

Keywords: Digital radiography, endodontics, root canal therapy

Introduction

Dental radiography plays a fundamental role in endodontic treatment. A comprehensive evaluation of root canal therapies cannot be accomplished without the help of imaging methods.1 Conventional periapical (PA) radiographs have always been the most common modality for this purpose.2 In addition, digital PA radiography is a newer technique with several advantages over the conventional one. Lower radiation dose, less time-consuming nature, the ability to process, modify, save and transfer the images and elimination of developing procedures are among these advantages, which have made the digital modality a favorable choice for endodontic therapy. However, the conventional technique has better spatial resolution and is less expensive.3-6

Precise root canal working length (WL) determination is an essential prerequisite for a successful root canal therapy (RCT). Endodontic mishaps such as apical perforation, improper cleaning and shaping or over-/under-filling may occur following an inaccurate WL determination, which can lead to RCT failure.7,8

There are different methods available for estimating the length of root canal such as tactile sensation, assessment of preoperative radiographs and electronic apex finders. The PA radiography, if used properly, can be a reliable and exquisite method for WL determination.,9 A parallel PA radiograph with an optimum contrast can readily illustrate the WL.10 The most common radiographs are conventional E- and F-speed films, alongside digital CCD (Charge Coupled Device) and PSP (Photo-Stimulable Phosphor Plates) images. Since in direct digital radiographic techniques such as CCD, the image is immediately observable after exposure, CCD seems to be more appropriate than PSP for endodontic therapies. However, the thickness and rigidity of CCD sensors limit their utility, whereas PSPs are more convenient and acceptable by patients.6,11,12 Therefore, choosing the best method could be controversial.

The diagnostic accuracy of conventional and digital PA radiographs has widely been investigated. Mostly, digital methods are authentic modalities with no significant difference from conventional ones.7,13,14

With the vast increase in the use of conventional and digital PA radiographs in endodontic treatments, we decided to compare the accuracy of the two approaches in root canal WL determination in an in vitro research.

Methods

In this in vitro study, 50 permanent and mature single-rooted maxillary central incisors were selected and investigated clinically and radiographically to be free of any fracture, internal/external resorption, extreme calcification, dilaceration or significant root curvature. The apical foramina of all these single-canal teeth were almost coincident with their anatomic apices.

After cleaning and disinfecting (5.25% sodium hypochlorite, Kimidaru, Iran), the access cavity was prepared (round and fissure burs, D&Z, Sydney, Australia) and the teeth were numbered consecutively. In order to obtain the actual WL (gold standard) a #15 K-file (Mani, Nakaakutsu, Japan) was placed tip-to-tip with the apical foramen. This was confirmed by a magnifying loupe. The rubber stop was carefully positioned to the reference point (incisal edge) and the file was removed from the root canal. The distance between the stopper and the file tip was measured with a caliper with an accuracy of 0.1mm (Mitutoyo, Tokyo, Japan) by an endodontist and a maxillofacial radiologist and then 1 mm was subtracted from that point to establish the actual WL (Gold1). The file was then placed back in the canal and fixed in place with Cavit dressing (Altstätten, Switzerland). A week later, this procedure was repeated on all the samples randomly (Gold2).2,4,15

Each tooth was mounted securely in the maxillary central incisor socket of a dry skull, which was fixed on a flat panel by plaster. This panel had a specified constant position and the radiography tube head was immobile during the entire imaging process. Each tooth was embedded in bone to the depth of the cementoenamel junction. In order to simulate soft tissue, the labial and palatal aspects of the maxilla were covered with eight layers and one layer of base plate wax at a consistent thickness, respectively.16-18 A Rinn film holder (Kerr, Germany) was fixed to the x-ray tube cone at a 5-cm distance from the embedded teeth (Figure 1).

Figure 1. Image capturing from the embedded tooth in the mounted skull.

The conventional radiographs were obtained with #2 E- and F-Speed intraoral films (Eastman-Kodak Co. Rochester, NY, USA) using a Prostyle dental x-ray machine (Planmeca, Helsinki, Finland) operated at 63 kVp, 8 mA and 0.25 s. All the films were developed with an automatic film processor (Hope Dental Max, Hope X-ray Products, USA) at the same time and temperature. The CCD sensor (De Gotzen, Italy) and PSP plate (Soredex, Helsinki, Finland) were exposed at 60 kVp, 7 mA and 0.06 s. The radiographs were all captured using the parallel technique.

On the conventional radiographs (E- and F-speed films), the distance between the incisal edge and file tip was measured with a caliper by an endodontist and a maxillofacial radiologist (E1, F1). This procedure was repeated a week later (E2, F2). The same process was carried out on CCD and PSP images using SOPRO® (Sopro Imaging, Sopro SA, La Ciotat, France) and Digora® (Digora for Windows 2.0, Soredex Medical Systems, Helsinki, Finland) software programs, respectively (CCD1, CCD2; PSP1, PSP2) (Figure 2). The measurements were all made in a semi-dark room under constant observational conditions. No visual enhancement was performed on digital images. These values were further assessed by means of statistical tests.

Figure 2. Measuring the working length in a CCD image using SOPRO software.

In order to ensure the consistency of the measuring accuracy of caliper and digital software programs’ measuring tools, a conventional, CCD and PSP image were obtained from an endodontic gauge‏.The same length was measured on the three images and no difference was noticed.

Statistical analysis

The correlation between the observers was assessed using Pearson's correlation test. Kolmogorov–Smirnov test showed the normality of data distribution. One-way ANOVA was used to compare the measuring accuracy of conventional and digital radiographs. Data were analyzed with SPSS 18 (SPSS Inc., Chicago, IL, USA) and the significance level was set to P<0.05.

Results

There was a significant correlation between the observers in conventional and digital radiographic techniques (P<0.001). The mean data obtained at the first and second week were considered as reference values for analysis (Table 1).

Table 1. The mean measured working length values
Group N Mean (Std. deviation)
Gold 1 50 21.57(2.32)
Gold 2 50 21.47(2.22)
E 1 50 21.88(2.29)
E 2 50 21.87(2.25)
F 1 50 22.02(2.27)
F 2 50 21.69(2.27)
CCD 1 50 21.59(2.22)
CCD 2 50 21.56(2.17)
PSP 1 50 21.69(2.23)
PSP 2 50 21.59(2.17)
Gold=Gold standard; E=E-speed film; F=F-speed film

In order to compare the measuring accuracy of different radiographs, the mean WL values (Gold, E, F, CCD and PSP) were compared. No significant difference was observed between the conventional, digital and gold standard values (P=0.828).

Discussion

Accurate root canal WL determination, a basic step in endodontic therapy, is accomplished by various techniques, using radiography, apex finders or tactile sensation.3,19,20 In this context, digital and conventional radiographs have been compared in vitro. The results of the present study showed no significant difference between the accuracy of the two techniques and the actual measurements.

In an in vitro and in vivo study which compared digital radiography (radiovisiography) with conventional method (D-Speed film), no significant difference was observed in the estimated canal lengths using the two modalities.21 Shearer et al22 reported no significant difference between conventional films and enhanced radiovisiography in estimating the root canal length. In an in vitro comparison of premolar root lengths, CCD, PSP and E-speed films exhibited similar measurement accuracy,23 consistent with the results of the present study.

A majority of studies have compared the two digital and conventional imaging systems in measuring the WL of curved canals. Burger et al24 evaluated direct digital radiography (DDR) versus conventional radiography for determining root canal length in curved canals. Although both techniques were deficient in measuring the true root canal length, there was no significant difference between the conventional and digital radiographs. Another in vitro study on 30 curved mesiobuccal root canals in molars compared conventional and PSP images in estimating the root canal length. The measurements for determining WL of curved canals obtained on either conventional or digital images were similar.2 Another research on 20 mesiobuccal canals from maxillary molars with moderate-severe curvatures and 20 canals form anterior teeth yielded the same results. Five radiographs were taken for each canal. Three conventional radiographies were obtained by using different processing techniques: Manual, automatic, and monobath solution. The two other digital images were captured using CCD and PSP receptors. The results suggested that the accuracy of digital and conventional radiographic techniques was comparable.25 However, some authors believe that digital systems are more accurate in canals with curvatures >25 degrees.26 The discrepancy might be due to the different measuring methods used. Some studies have proved that conventional film is superior in WL measurement when compared to digital systems.27-30 This is mostly due to the old digital systems. On the other hand, some other studies have shown digital systems to be superior to conventional films.31,32 The numerous features available for image enhancement in digital systems might justify these findings. Computerized enhancement is more useful when detecting file tip positions, especially of small sizes. This is one reason for suggesting digital radiography in endodontic practice, albeit it mostly shows similar accuracy with the conventional one. Furthermore, digital radiography has lower radiation dose and is less time-consuming.5

Apex-locators are also compared with radiography in determining the root canal WL. Some researchers claim that the new generation apex finders could estimate the WL better than radiography.33 However, it totally depends on the accuracy of the technique used.

Conclusion

The measurement accuracy of CCD, PSP and conventional films in WL determination exhibited no difference between these techniques. However, because of the advantages of digital systems (e.g. radiation dose reduction or image enhancement), it is preferable to the conventional one.

Acknowledgments

The authors would like to express their gratitude to Dr. Entezari and Dr. Darabinasab for sharing their pearls of wisdom and efforts with us during the course of this research.

Authors’ contributions

YKMM collected data and interpreted them. EH analyzed data analysis and drafted and revised the manuscript. Both authors have read and approved the final manuscript.

Funding

Rafsanjan School of Dentistry financially supported the present study.

Competing interests

The authors declare no competing interests with regards to the authorship and/or publication of this article.

Ethics approval

The study was approved by Rafsanjan University of Medical Sciences.

References

  1. Mohtavipour ST, Dalili Z, Azar NG. Direct digital radiography versus conventional radiography for estimation of canal length in curved canals. Imaging Sci Dent 2011;41(1):7-10.doi: 10.5624/isd.2011.41.1.7. [Crossref]
  2. Brito-Júnior M, Santos LA, Baleeiro ÉN, Pêgo MM, Eleutério NB, Camilo CC. Linear measurements to determine working length of curved canals with fine files: conventional versus digital radiography. J Oral Sci 2009;51(4):559-64. doi: 10.2334/josnusd.51.559. [Crossref]
  3. Pour DG, Razmi H, Maralani SJ, Zeighami S. New software: comparison between three software programs for root canal length measurement. Dentomaxillofac Radiol 2008;37(4):228-31. doi: 10.1259/dmfr/16010494. [Crossref]
  4. Woolhiser GA, Brand JW, Hoen MM, Geist JR, Pikula AA, Pink FE. Accuracy of film-based, digital, and enhanced digital images for endodontic length determination. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2005;99(4):499-504. doi: 10.1016/j.tripleo.2004.07.024. [Crossref]
  5. Wenzel A, Møystad A. Work flow with digital intraoral radiography: a systematic review. Acta Odontol Scand 2010;68(2):106-14. doi: 10.3109/00016350903514426. [Crossref]
  6. Van der Stelt PF. Better imaging: the advantages of digital radiography. J Am Dent Assoc 2008;139:S7-S13.doi:10.14219/jada.archive.2008.0357. [Crossref]
  7. Mentes A, Gencoglu N. Canal length evaluation of curved canals by direct digital or conventional radiography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2002;93(1):88-91.doi:10.1067/moe.2002.119466. [Crossref]
  8. Martín-Micó M, Forner-Navarro L, Almenar-García A. Modification of the working length after rotary instrumentation. A comparative study of four systems. Med Oral Patol Oral Cir Bucal 2009;14(3):E153-7.
  9. Gordon M, Chandler N. Electronic apex locators. Int Endontic J 2004;37(7):425-37.doi:10.1111/j.1365-2591.2004.00835.x. [Crossref]
  10. Sheaffer JC, Eleazer PD, Scheetz JP, Clark SJ, Farman AG. Endodontic measurement accuracy and perceived radiograph quality: effects of film speed and density. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003;96(4):441-8.doi:10.1016/s1079-2104(03)00035-0. [Crossref]
  11. Almeida SMd, Oliveira AEFd, Ferreira RI, Bóscolo FN. Image quality in digital radiographic systems. Braz Dent J 2003;14(2):136-71.doi:10.1590/S0103-64402003000200012. [Crossref]
  12. Bóscolo FN, Oliveira AE, Almeida SMd, Haiter C, Haiter Neto F. Clinical study of the sensitivity and dynamic range of three digital systems, E-speed film and digitized film. Braz Dent J 2001;12(3):191-5.
  13. Schulze D, Rother U, Fuhrmann AW, Tietke M. A comparison of two intraoral CCD sensor systems in terms of image quality and interobserver agreement. Int J Comput Dent 2003;6(2):141-50.
  14. Benediktsdottir I, Hintze H, Petersen JK, Wenzel A. Image quality of two solid-state and three photostimulable phosphor plate digital panoramic systems, and treatment planning of mandibular third molar removal. Dentomaxillofac Radiol 2003;32(1):39-44.doi:10.1259/dmfr/25838744. [Crossref]
  15. Soundappan S, Sundaramurthy JL, Raghu S, Natanasabapathy V. Biodentine versus Mineral Trioxide Aggregate versus Intermediate Restorative Material for Retrograde Root End Filling: An Invitro Study. J Dent (Tehran) 2014;11(2):143.
  16. Blattner TC, George N, Lee CC, Kumar V, Yelton CD. Efficacy of cone-beam computed tomography as a modality to accurately identify the presence of second mesiobuccal canals in maxillary first and second molars: a pilot study. J Endod 2010;36(5):867-70.doi:10.1016/j.joen.2009.12.023. [Crossref]
  17. Salemi F, Saati S, Falah-Kooshki S. A comparative study of the conventional and digital intraoral radiography methods for root canal length measurement. Braz Dent Sci 2014;17(4):34-9.doi:10.14295/bds.2014.v17i4.997. [Crossref]
  18. Akdeniz B, Soğur E. An ex vivo comparison of conventional and digital radiography for perceived image quality of root fillings. Int Endontic J 2005;38(6):397-401.doi:10.1111/j.1365-2591.2005.00958.x. [Crossref]
  19. Hoer D, Attin T. The accuracy of electronic working length determination. Int Endod J 2004;37(2):125-31.doi:10.1111/j.0143-2885.2004.00764.x. [Crossref]
  20. Jafarzadeh H, Abbott PV. Ledge formation: review of a great challenge in endodontics. J Endod 2007;33(10):1155-62.doi:10.1016/j.joen.2007.07.015. [Crossref]
  21. Ong E, Ford TP. Comparison of radiovisiography with radiographic film in root length determination. Int Endontic J 1995;28(1):25-9.doi:10.1111/j.1365-2591.1995.tb00152.x. [Crossref]
  22. Shearer AC, Horner K, Wilson NH. Radiovisiography for length estimation in root canal treatment: an in-vitro comparison with conventional radiography.Int Endod J 1991;24(5):233-9.
  23. Velders XL, Sanderink GC, van der Stelt PF. Dose reduction of two digital sensor systems measuring file lengths. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996;81(5):607-12.doi:10.1016/s1079-2104(96)80057-6. [Crossref]
  24. Burger CL, Mork TO, Hutter JW, Nicoll B. Direct digital radiography versus conventional radiography for estimation of canal length in curved canals. J Endod 1999;25(4):260-3.doi:10.1016/s0099-2399(99)80155-1. [Crossref]
  25. Farida A, Maryam E, Ali M, Ehsan M, Sajad Y, Soraya K. A comparison between conventional and digital radiography in root canal working length determination. Indian J Dent Res 2013;24(2):229.doi:10.4103/0970-9290.116693. [Crossref]
  26. Ezoddini Ardakani F, Mohammadi Z. Comparison between the Accuracy of Digital and Conventional Radiographies for Evaluation of the Curved Canals Length. Dent Res J (Isfahan) 2008;3(2):1-10.
  27. Sanderink GC, Huiskens R, Van der Stelt P, Welander US, Stheeman SE. Image quality of direct digital intraoral x-ray sensors in assessing root canal length: the RadioVisioGraphy, Visualix/VIXA, Sens-A-Ray, and Flash Dent systems compared with Ektaspeed films. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1994;78(1):125-32.doi:10.1016/0030-4220(94)90128-7. [Crossref]
  28. Ellingsen MA, Harrington GW, Hollender LG, Odont D. Radiovisiography versus conventional radiography for detection of small instruments in endodontic length determination. Part 1. In vitro evaluation. J Endod 1995;21(6):326-31.doi:10.1016/s0099-2399(06)81010-1. [Crossref]
  29. Lozano A, Forner L, Llena C. In vitro comparison of root‐canal measurements with conventional and digital radiology. Int Endontic J 2002;35(6):542-50.doi:10.1046/j.1365-2591.2002.00535.x. [Crossref]
  30. Hedrick RT, Dove SB, Peters DD, McDavid WD. Radiographic determination of canal length: direct digital radiography versus conventional radiography. J Endod 1994;20(7):320-6.doi:10.1016/s0099-2399(06)80093-2. [Crossref]
  31. Radel RT, Goodell GG, McClanahan SB, Cohen ME. In vitro radiographic determination of distances from working length files to root ends comparing Kodak RVG 6000, Schick CDR, and Kodak insight film. J Endod 2006;32(6):566-8.doi:10.1016/j.joen.2006.01.001. [Crossref]
  32. Nair MK, Nair UP. Digital and advanced imaging in endodontics: a review. J Endod 2007;33(1):1-6.doi:10.1016/j.joen.2006.08.013. [Crossref]
  33. Tampelini FG, Coelho MS, Rios MD, Fontana CE, Rocha DG, Pinheiro SL, et al. In vivo assessment of accuracy of Propex II, Root ZX II, and radiographic measurements for location of the major foramen.Restor Dent Endod 2017;42:e31.doi:10.5395/rde.2017.42.e31. [Crossref]
First published online: 20 Sep 2017
EndNote EndNote

(Enw Format - Win & Mac)

BibTeX BibTeX

(Bib Format - Win & Mac)

Bookends Bookends

(Ris Format - Mac only)

EasyBib EasyBib

(Ris Format - Win & Mac)

Medlars Medlars

(Txt Format - Win & Mac)

Mendeley Web Mendeley Web
Mendeley Mendeley

(Ris Format - Win & Mac)

Papers Papers

(Ris Format - Win & Mac)

ProCite ProCite

(Ris Format - Win & Mac)

Reference Manager Reference Manager

(Ris Format - Win only)

Refworks Refworks

(Refworks Format - Win & Mac)

Zotero Zotero

(Ris Format - FireFox Plugin)

Abstract View: 1685
PDF Download: 1700
Full Text View: 1003